DEZEGA CheckUp breathing equipment test set checks the parameters of compressed air and compressed oxygen breathing apparatuses, as well as self-contained panoramic masks and resuscitators. The test set should be used by the departments that operate or service BA. The test set complies with Directive 2014/30/EU, Directive 2014/35/ЕU, EN 61326-1:2013.

CATALOGUE
Описание
Description

DEZEGA CheckUp will check :

  • self-contained breathing apparatuses compliant with EN 145:1997, DSTU EN 145, DSTU 3856, GOST R 12.4.253, GOST R 53256.
  • self-contained panoramic masks compliant with EN 136:1998, DSTU EN 136, GOST R 53257 or equivalent;
  • resuscitators of the GS-10 type;
  • separate components of the breathing equipment.
Характеристики
Features

Pressure measuring range (positive-negative), kPa

from - 2,8 to 6,8

Limits of permissible overpressure-dilution measurement error, % of range measurements

 

 

±0,5

Range of indications of flow through the oxygen channel, dm3/min

from 0 to 6

Range of measurements of flow through the oxygen channel, dm3/min

from 1 to 6

Limits of permissible reduced error per channel oxygen, % of the upper limit of measurement

in the range from 1 to 3.0 dm3/min
in the range above 3.0 to 6.0 dm3/min

 

±2,0

±3,0

Range of indications of an expense on the air channel, dm3/min

from 0 to 150

Range of measurements of an expense on the air channel, dm3/min

from 10 to 150

Limits of permissible reduced error per channel airflow, % of the upper limit of measurement

±5,0

 

Operation mode

Duration of one cycle, min

  • Break between cycles, min, not less than
  • Continuous operation time in multiple cyclic mode, h (per day), not more than

Cyclical

from 5 to 30

5

7

Power supply parameters:

  • AC voltage, V
  • AC frequency, Hz
  • power consumption, V∙A, not more than

 

220±15%

50±5

230

Overall dimensions, mm, not more than

  • height
  • length
  • width

 

220

460

480

Weight (without adapters), kg, not more than

15

Operating conditions:

  • ambient temperature, °С
  • relative humidity (without condensation), %
  • Atmospheric pressure, kPa

 

from +15 to +25

from 30 to 80

from 84 to 106,7

Service life from the date of manufacture, years

9

Mean time to failure, h

16240

Документы
Files
User manual
User manual DEZEGA CheckUp
User manual describes the construction and requirements for operation and service of DEZEGA CheckUp. Languages: EN, UK.
Download
Certificate
Certificate
EU-type examination certificate.
Download
Declaration of conformity CheckUp
Declaration of conformity with the requirements of the Technical Regulations of Ukraine EMC (Resolution of the Cabinet of Ministers of Ukraine #1077).
Download
Certificate
Certificate of conformity with requirements of the Technical Regulations of Ukraine EMC (Resolution of the Cabinet of Ministers of Ukraine #1077).
Download